Product Specifications PICK-O2 All product specifications are applicable at standard conditions: 1013 hPa, 25°C dry ambient air. Measuring principle: Galvanic oxygen sensor Measurement Range: 0 - 99 % vol oxygen Display resolution: 1 % vol Signal response: < 1 sec. Connections: Instrument: M16 x 1 Flowdiverter: sealed plug-in Accuracy: ± 3 % of the reading value (calibrated with dry air) Interference: less than 0.1 % O2 response to: 15.0 % CO2 balance N2 10.0 % CO balance N2 3000 ppm NO balance N2 3000 ppm C3H8 balance N2 500 ppm H2S balance N2 500 ppm SO2 balance N2 1000 ppm Benzene balance N2 Operating Temperature: 0 - 50°C Optimal Storage Temperature: 5 - 25°C Maximum Storage Temperature: -5 - 60°C Instrument lifetime: ³ 2 years Weight: approximately 80 grams Dimensions: H = 74 mm (2,9"), W = 47 mm (1,85"), D = 54 mm (2,1"), Influence of Pressure and Moisture: corresponds to partial pressure-change of O2 Specifications are subject to change without prior notice.